HYDRO-SET
User Manual Version 1.49 (English) © 2001-2009 HYDROMETER GmbH / Ansbach / Germany
-
What is the HYDRO-SET Software anyway? Versions System Requirements Software Installation Customer Support HYDRO-SET Software In General 4 Main Window Main Window: Communication 1 Phone Number Administration Main Window: Communication 2 Main Window: User and Status Elements Settings SHARKY-HEAT…
-
Display 1
Display 2-6 Tariff Settings Calibration Telegram SHARKY SW29 Energy Meter 4 (SW 29) Current Values (SW 29) Reading Dates Values 1 (SW 29) Reading Dates Values 2 (SW 29) Maximum Values (SW 29) Average Values (SW 29) Leakage (SW 29) Monthly Log (SW 29) EEPROM (SW 29) EEPROM Configuration (SW 29)…
-
FLYPPER II and SCAMPY Water Meters 4 Current Values Reading Dates Values Monthly Log EEPROM Settings FLYPPER III and SCAMPY II Water Meters 4 Current Values Reading Dates Values Monthly Log Settings FLYPPER IV Water Meter 4 Current Values Reading Dates Values Monthly Log Errors Settings ENERGIE-INT…
-
(INT7) Average Values
(INT7) Leakage (INT7) Monthly Log (INT7) EEPROM (INT7) EEPROM Configuration (INT7) Errors (INT7) Display 1 (INT7) Display extended (INT7) Tariff (INT7) Impulses (INT7) Settings (INT7) Calibration (INT7) Telegram HYDRUS water meter 4 (HYDRUS) Current Values (HYDRUS) Reading Date Values (HYDRUS) Log…
-
© HYDROMETER GmbH Ansbach / Germany / 2001 - 2008
The name HYDRO-SET®, the HYDRO-SET software and this manual are protected by copyright laws. Copying, translating, transferring to other media like microfiches and other electromagnetic or optical storage media without the written permission of HYDROMETER is prohibited. Trademarks or registered…
-
What is the HYDRO-SET Software anyway?
The HYDRO-SET software is used for reading, configuration and with some limitations also for (post-) calibration of heat meter and water meter products. Currently (September 2008) the devices: • SHARKY-HEAT (BR770), SHARKY-VMC (BR471), Cétas ultrason, F95 • SHARKY-HEAT 130 °C (BR772) , SHARKY-VMC…
-
Versions
There are two different versions of this software. HYDRO-SET: With the freely available standard version of HYDRO-SET you are able to read all meter values and you may program several user specific meter parameters (e.g. device address, reading dates). HYDRO-SET Professional: The professional…
-
System Requirements
The HYDRO-SET software runs on a Pentium class PC computer with at least 256 MByte of free memory, Windows® XP and approx. 40 MByte free hard disk space. Additionally, a free serial interface port must be available. For connecting a heat meter or a water meter, an opto transceiver (optical reading)…
-
таблицы и схемы — в просмотре
-
HYDRO-SET Software In General
Main Window The HYDRO-SET software follows the Windows® standards for PC software. Entry fields with a white background may be altered, entry fields with a yellow background are fixed and are used for displaying values only. If you own the professional version the text “professional” appears in the…
-
After the start of HYDRO-SET the dialog box above is shown. It is split in two parts
• Communication Here you select how to communicate with your device and how the connection is established. • Device specific property pages The property pages are disabled after the program start (gray field in the lower part of the screen). Depending on the device connected, its parameters and…
-
Select "Optical Communication (According To Settings)" if you are using an opto transceiver to
communicate with the M-Bus device or the M-Bus repeater (HYDRO-CENTER). You have to set the device type of the opto transceiver at the settings dialog (Settings). Depending on the device type the entry field for the baudrate setting is visible or not. The standard baudrate for optical communication…
-
Edit
Select an entry from the list and press "Edit". The same dialog as for "New" appears and you are able to edit an existing phone number. Delete: Select an entry from the list and press "Delete". The selected entry is deleted from the list. For selecting a phone number you have to single-click an…
-
Main Window: Communication 2
There are three different ways to address a device. You may select them as described beneath. However, these settings are not to be mixed up with the setting of the connection type (Main Window: Communication 1). Select "M-Bus Point-to-Point (Addr. 254)" if you are communicating using a opto…
-
Main Window: User and Status Elements
Status Line and Progress Bar: The entry field with the yellow background displays the current program state . The progress bar beneath shows the progress of the currently ongoing operation (e.g. reading a device). 1 Read: You start a device read by pressing "Read". During the reading process the…
-
(HYDRUS) Current Values
The following parameters are shown in the top line of the display: Meter type: Hydrus with Q3 and interfaces M-Bus primary address: 124 M-Bus secondary address / device address: 12345678 Note: To change the device address, you must first activate the input field in the settings. Manufacturer’s ID…
-
The current values in the meter are shown below
Set Date+Time: For setting the date and time to the given value. Type in the date or click Set Date+Time to display the calendar input window. Synchronize With PC: For adjusting the date and time to the current time of the computer. Note: A mouse click on the lettering after the date input field…
-
(HYDRUS) Reading Date Values
The values for reading date 1 are shown. pY: previous year Reading Date 1 (next): For setting the next reading date for reading date 1. Type in the reading date or click Date to display the calendar input window. Notes: The reading date values always refer to the end of the day, i.e. to 23.59 hours…
-
(HYDRUS) Log Storage
This window shows the contents of the log storage memory. The log storage is not read automatically in a standard request, but must be read explicitly using the relevant button. The meter can store maximum 32 data records. These comprise: - Maximum flow - Minimum flow - Total volume - Return volume…
-
(HYDRUS) Error
A list of the last 31 events / errors is shown. The error log is not read automatically in a standard request, but must be read explicitly using the relevant button. Only available in the professional version! Delete Error Log: The error log is deleted completely. Save Values: Saves the errors shown…
-
(HYDRUS) Display
This tab is for programming the selection of fields and the sequence of displayed values for the meter display. The left list contains the possible values for the display and the list on the right shows the previously defined display values. The display configuration is set by selecting values…
-
(HYDRUS) Settings
Shows internal parameters of the meter that can be used for tasks such as fault diagnosis. To change parameters on this tab, the calibration button inside the meter must always be pressed. Pressing the calibration button in the meter changes the programming protection level from normal mode to…
-
The maximum output frequency of the constant 125 ms long pulse is 4 Hz. An open collector output is
used and the pulse output is obtained by connecting the applied voltage to earth for 125 ms. Total volume: The total volume in conjunction with the direction information (pulse output 2) represents in principle the total of the forward and return flows at any time. No return pulses are buffered and…
-
(HYDRUS) Calibration
This tab shows other internal parameters of the meter, which can be used for tasks such as fault diagnosis. To change parameters on this tab, the calibration button inside the meter must always be pressed. Pressing the calibration button in the meter changes the programming protection level from…
-
Offset Calibration On
(HYDRUS) Telegram The meter offers the user the possibility of compiling a customer-specific M-Bus telegram of his choice. This customer-specific M-Bus telegram is also used as a radio telegram if the meter is used with a Real Data Radio Module, i.e. the values to be sent over Real Data Radio are…
-
• Each value can only be used once in the customer telegram
• Changes to the customer telegram (right list) are not transferred to the meter until the Set Customer Telegram button has been pressed. Compiling the values in the right list is not sufficient to activate the customer telegram. A different telegram can be set as standard if requested by the…
-
How Do I …?). Before you are able to see or set any device parameter you have to read the device at
least once. ALT+1 is the hotkey for "Read". 2 Write: Using "Write" you are programming the device with a complete set of predefined parameters. "Write" is enabled only if you have loaded a parameter set for the respective device. ALT+2 is the hotkey for "Write". 3 Load: Using "Load" you can display…
-
Click the program icon in the upper left corner of the main window and select "Settings" from the menu. The dialog beneath appears: Table Of Contents You start a device read by pressing "Read". During the reading process the program establishes a connection to the device, checks the device type and…
-
Language
Select the language of the program texts. Currently the HYDRO-SET software is available in English, French, German, Italian and Danish. Phone Connection Timeout: Enter the time in seconds after which an attempt to establish a phone connection is recognized as failure. The default value is between 40…
-
Optical Communication
Select the device type of the optical transceiver you are using. This setting is only used if you have chosen “Optical Communication” on the main screen (see Main Window: Communication 1). For a compact V communication you have to select IrDA Optotransceiver HY-Group. Multiplex: Enables a multiplex…
-
SHARKY-HEAT (BR770) / SHARKYVMC (BR471)
Current Values The first line of entry fields indicates: Device Type and Class: SHARKY-HEAT qp 1.5 M-Bus Primary Address: 8 M-Bus Secondary Address / Device Address: 26718749 Note: If you want to change the device address you have to enable the respective entry field at the Settings dialog.…
-
Date+Time set: Date and time are set to the entered value. Enter the new date
manually or click on the description Date+Time Set on the right hand side of the date entry field and use the upcoming calendar to enter the date. Synchronize With PC: Date and time are synchronized with the PC clock. Note: By clicking the description on the right hand side of the date entry field…
-
The latest reading date values (Energy Reading Date 1 and Energy Reading Date 2) and, additionally,
their time points (Reading Date 1 and Reading Date 2) are displayed. Reading Date 1 (next): Setting of next time point for reading date 1. Enter the new date manually or click on the description Date+Time Set on the right hand side of the date entry field and use the upcoming calendar to enter the…
-
Clear All: All maximum values and their respective time points are cleared
Integration Time: The integration time in minutes for calculating the maximum values. Possible values are 6, 15, 30 and 60 minutes. Table Of Contents SHARKY-HEAT (BR770) / SHARKY-VMC (BR471) 36
-
Depending on the setting the maximum monthly power or the end of month values of the energy for 12
months are listed. Only available with the professional version ! Delete All Memory Values: The monthly log memory, all historic values and all current values are cleared (energy counter, volume counter, hours run counter, reading date values, maximum values, monthly log). This function can only be…
-
таблицы и схемы — в просмотре
-
This list displays the content of the EEPROM data logger memory of the meter. If data logging with
maximum values is enabled, the meter may store up to 494 time points with different values, if data logging without maximum values is enabled, the meter may store up to 988 time points. Due to the large size of the EEPROM data logger memory, these values are not loaded automatically during a…
-
The default value for the storage interval is 24 hours
Read EEPROM Memory: Reads out the EEPROM data logger memory. Depending on the number of time points stored this process may take up to 10 minutes (2400 baud). Therefore, it is possible to limit the number of time points to read. Table Of Contents SHARKY-HEAT (BR770) / SHARKY-VMC (BR471) 40
-
Events
A list with the latest 15 events / errors is displayed. Clear All: The event memory is cleared, completely. Table Of Contents SHARKY-HEAT (BR770) / SHARKY-VMC (BR471) 41
-
Here you are able to select the display sequence of the device. The display has got three loops. By
default loop 1 is displayed. By pressing the user button on the device you cycle through the values of loop 1. By pressing the user button for a longer time the display switches to the next loop and so on. The display of the device is organised in blocks. You are only able to enable or disable a…
-
Internal parameters of the device, which may be used for trouble shooting, are displayed. You have to press the protection key inside the device to alter any of the parameters on this page. Type of Meter: Indicates the type of the meter (e.g. heat meter, volume counting unit, …) Only available with…
-
Meter Energy Unit (Display): The default display unit (Wh or J) and the number of
decimal digits is selected here. Parameter (Display): Internal parameter for the display setting Only available with the professional version ! Installation: The device may be installed at the inlet or outlet (default) of a heating pipe system. Table Of Contents SHARKY-HEAT (BR770) / SHARKY-VMC…
-
Temperature Sensor: Indicates the type of the temperature sensor (Pt100, Pt500)
Restart Counter: Number of restarts of the internal meter firmware. Error State: Indicates the currently existing error state of the device (e.g. missing sensor or air in US path). Energy Pulse Output: Indicates if the device has got an energy pulse output or not. The pulse valence of this output is…
-
таблицы и схемы — в просмотре
-
Internal parameters of the device, which may be used for trouble shooting, are displayed. You have to press the protection key inside the device to alter any of the parameters on this page. Serial Number: Apart from the M-Bus device ID / secondary address the device has got an internal serial…
-
таблицы и схемы — в просмотре
-
Parameter KA / KB: Correction values A and B, two parameter for the flow measurement curve
Only available with the professional version ! Flow Calibration: You may decrease or increase the internal parameters KA and KB which are responsible for the volume flow error curve. E.g. if the device measures +1% volume flow you may set the adjustment to -1% to compensate the error. NOTE: The…
-
таблицы и схемы — в просмотре
-
SHARKY-HEAT 130 (BR772) Heat Meter
Current Values The first line of entry fields indicates: Device Type and Class: SHARKY-HEAT 130 qp 1.5 M-Bus Primary Address: 2 M-Bus Secondary Address / Device Address: 12345678 Note: If you want to change the device address you have to enable the respective entry field at the Settings dialog.…
-
manually or click on the description Set Date+Time on the right hand side of the date entry field and use the upcoming calendar to enter the date. Synchronize With PC: Date and time are synchronized with the PC clock. Note: By clicking the description on the right hand side of the date entry field…
-
The latest reading date values of Energy, Energy Tariff 1 and Energy Tariff 2 and, additionally, their time points (Reading Date 1 and Reading Date 2) are displayed. Reading Date 1 (next): Setting of next time point for reading date 1. Enter the new date manually or click on the description Date on…
-
Integration Time: The integration time in minutes for calculating the maximum values
Possible values are 6, 15, 30 and 60 minutes. Table Of Contents SHARKY-HEAT 130 (BR772) Heat Meter 54
-
Different end of month values of the last 24 months are displayed. The monthly log is not loaded
automatically during a standard request, instead, the user has to request it separately. Save Values: Stores the list of end of month values into a text file. Read Monthly Log: Reads the monthly log memory of the meter. This is not performed automatically during a standard request. Table Of Contents…
-
The content of the EEPROM data logger memory is displayed. This meter may store up to 468 time
points with different values. Due to the large size of the EEPROM data logger memory it is not loaded automatically during a standard request, instead, the user has to request it separately. Please note that the refreshing of the list, depending on the number of values read, may take from several…
-
on the number of time points stored this process may take up to 5 minutes (2400 baud). Therefore, it is possible to limit the number of time points to read. Table Of Contents SHARKY-HEAT 130 (BR772) Heat Meter 57
-
A list with the latest 31 events / errors is displayed. The error log is not automatically loaded during a standard request, instead, it has to be requested separately. Only available with the professional version ! Reset Error Log: The error log is deleted completely.. Save Values: Stores the…
-
Here the user may enable or disable certain values in the display of the meter. By default the meter
displays display loop 1. By pressing the user button on the device you cycle through the values of display loop 1. By pressing the user button for a longer time the display switches to display loop 2 and so on. Set Values For Display: The values to display shown in the list are programmed into the…
-
The meter has got two programmable tariffs for the energy value. The energy values of the two tariffs are also stored at reading date 1 and reading date 2. Tariff 1 Limit: Set type and limit of tariff 1. Tariff 2 Limit: Set type and limit of tariff 2. In addition to tariff 1 tariff 2 offers the…
-
Internal parameters of the device, which may be used for trouble shooting, are displayed. You have to press the protection key inside the device to alter any of the parameters on this page (with the exception of the impulse settings). After pressing the protection key inside the device the…
-
pipe system
Temperature Sensor: Indicates the type of the temperature sensor (Pt100, Pt500). Errors: Indicates the number of days with errors (since installation of the meter or since last reset of the error log). Error State: Indicates the currently existing error state of the device (e.g. missing sensor or…
-
Impulse Output 1 / 2: Sets the type of the impulse outputs 1 and 2. In general the
meter may output an energy or volume proportional impulse or a state output which will be set if one of the following conditions is met: • Condition Tariff 1: Tariff 1 is active • Condition Tariff 2: Tariff 2 is active • Condition Error E: is set if one or more of the following errors are found: •…
-
Internal parameters of the device, which may be used for trouble shooting, are displayed. You have to press the protection key inside the device to alter any of the parameters on this page. After pressing the protection key inside the device the protection level is switched from normal mode to…
-
not working properly
E.g. if the device measures +1% volume flow you may set the adjustment to -1% to compensate the error. Only available with the professional version ! Start Offset Calibration: The automatic offset calibration process for the ultrasonic measuring chamber is started. The process takes about 1 minute…
-
SHARKY-HEAT 130 (BR772) Heat Meter
-
SHARKY (BR773) Energy Meter
Note This device is available in a version with leakage detection and in a version without leakage detection. The device with leakage detection has got some more settings and options (leakage, display, pulse inputs, customer telegram), which are not available for the other version. Current Values…
-
manually or click on the description Set Date+Time on the right hand side of the date entry field and use the upcoming calendar to enter the date. Synchronize With PC: Date and time are synchronized with the PC clock. Note: If the meter is in normal mode, it is not possible but to set the time. The…
-
Reading Dates Values 1
The values for reading date 1 are shown. LY: last year’s value T1: tariff 1 T2: tariff 2 Reading Date 1 (next): Setting of next time point for reading date 1. Enter the new date manually or click on the description Date on the right hand side of the date entry field and use the upcoming calendar to…
-
Reading Dates Values 2
The values for reading date 2 are shown. LY: last year’s value T1: tariff 1 T2: tariff 2 Reading Date 2 (next): Setting of next time point for reading date 2. Enter the new date manually or click on the description Date on the right hand side of the date entry field and use the upcoming calendar to…
-
Integration Time: The integration time in minutes for calculating the maximum values
Possible values are 6, 15, 30 and 60 minutes and 24 hours. 60 minutes is the default integration time. Table Of Contents SHARKY (BR773) Energy Meter 71
-
Leakage
This option page with leakage settings is not available but for devices with leakage option. Acknowledge Alarm: An eventual alarm is reset. Leakage Detection (Heat): The leakage detection for the heat circuit may be switched on or off. With the same option dialog the accuracy / sensitivity for the…
-
Integration Time: The integration time of the leakage detection is set with this option
Possible integration times are between 1 hour and 24 hours. Stop Time Input 1: Alternatively to the leakage detection for the heat circuit, the device may set an alarm if there is not a certain interval of time within the alarm integration time without an impulse on impulse input 1. Usually, a…
-
Different month values of the last 24 months are displayed. The monthly log is not loaded automatically during a standard request, instead, the user has to request it separately. Save Values: Stores the list of end of month values into a text file. Log Day: The day of the month, at which the month…
-
The content of the EEPROM data logger memory is displayed. This meter may store up to 440 time
points with different values. Due to the large size of the EEPROM data logger memory it is not loaded automatically during a standard request, instead, the user has to request it separately. Please note that the refreshing of the list, depending on the number of values read, may take from several…
-
The default value for the storage interval is 24 hours, EEPROM memory 2 is not used
Read EEPROM Memory: Reads out the EEPROM data logger memory. Depending on the number of time points stored this process may take up to 5 minutes (2400 baud). Therefore, it is possible to limit the number of time points to read. Read EEPROM Memory 2 : Reads out the EEPROM data logger memory 2. If the…
-
A list with the latest 31 events / errors is displayed. The error log is not automatically loaded during a standard request, instead, it has to be requested separately. Only available with the professional version ! Reset Error Log: The error log is deleted completely.. Save Values: Stores the…
-
Display 1
Here the user may set the sequence of the displayed values in display loop 1 of the meter. The list on the left hand side contains all possible display values, the list on the right hand side shows the currently programmed display sequence of display loop 1. By selecting values (multiple selections…
-
SHARKY (BR773) Energy Meter
-
Display 2-6
Here the user may enable or disable certain values in the display of the meter (display loop 2 – 6). Usually the meter displays the display loop 1, that is by pressing the user button on the meter the user cycles through the display values of display loop 1. The setting of display loop 1 is…
-
The meter has got two programmable tariffs for the energy value. The energy values of the two tariffs are also stored at reading date 1 and reading date 2. Clear Tariff 1: Clear values for tariff 1. Tariff 1 Limit: Set type and limit of tariff 1. For tariff type t08 and t09 an additional minimum…
-
possibility to logically AND tariff 1 and tariff 2
Table Of Contents SHARKY (BR773) Energy Meter 82
-
Internal parameters of the device, which may be used for trouble shooting, are displayed. You have to press the protection key inside the device to alter any of the parameters on this page (with the exception of the impulse settings). After pressing the protection key inside the device the…
-
pipe system
Temperature Sensor: Indicates the type of the temperature sensor (Pt100, Pt500). Nur in der Professionell Version verfügbar ! Errors: Indicates the number of days with errors (since installation of the meter or since last reset of the error log). For some devices with leakage detection the error day…
-
(without leakage detection)
Additionally, the current impulse input value can be reset to zero during programming the impulse input weight. This option is only available with devices not having leakage detection. With devices having leakage detection the meter count on the impulse inputs may be freely set. (with leakage…
-
Impulse Output 1 / 2: Sets the type of the impulse outputs 1 and 2. In general the
meter may output an energy or volume proportional impulse or a state output which will be set if one of the following conditions is met: • Energy: Pulse weight is the lowest digit of the energy display • Energy / 10: Pulse weight is the lowest digit of the energy display / 10 (i.e. Energy = 34,589…
-
• Leakage at pulse input 1
Pulse output is active if leakage at pulse input 1 is recognized • Leakage at pulse input 2: Pulse output is active if leakage at pulse input 2 is recognized • Leakage at pulse input 1 or 2: Pulse output is active if leakage at pulse input 1 or 2 is recognized • Deactivated: Pulse output is not…
-
Internal parameters of the device, which may be used for trouble shooting, are displayed. You have to press the protection key inside the device to alter any of the parameters on this page. After pressing the protection key inside the device the protection level is switched from normal mode to…
-
not working properly
E.g. if the device measures +1% volume flow you may set the adjustment to -1% to compensate the error. Only available with the professional version ! Start Offset Calibration: The automatic offset calibration process for the ultrasonic measuring chamber is started. The process takes about 1 minute…
-
Telegram
For this meter the user may assemble his own customer specific M-Bus telegram. For meters equipped with a real data radio module this customer telegram is also used as radio telegram, therefore, the user may set the values to be transmitted via real data radio. If the customer telegram is used as…
-
• All changes to the right hand side list (customer telegram) are not taken unless the user has
programmed them using the Set customer telegram button. The assembly of the values in the right hand side list is not sufficient for activating the customer telegram. • If the list on the right hand side is completely empty, the standard customer telegram is used: • [03] Energy • [04] Volume • [05]…
-
SHARKY SW29 Energy Meter
(SW 29) Current Values The first line of entry fields indicates: Device Type and Class: SHARKY qp 1.5 M-Bus Primary Address: 0 M-Bus Secondary Address / Device Address: 12345678 Note: If you want to change the device address you have to enable the respective entry field at the Settings dialog.…
-
manually or click on the description Set Date+Time on the right hand side of the date entry field and use the upcoming calendar to enter the date. Synchronize With PC: Date and time are synchronized with the PC clock. Note: By clicking the description on the right hand side of the date entry field…
-
(SW 29) Reading Dates Values 1
The values for reading date 1 are shown. LY: last year’s value T1: tariff 1 T2: tariff 2 Reading Date 1 (next): Setting of next time point for reading date 1. Enter the new date manually or click on the description Date on the right hand side of the date entry field and use the upcoming calendar to…
-
(SW 29) Reading Dates Values 2
The values for reading date 2 are shown. LY: last year’s value T1: tariff 1 T2: tariff 2 Reading Date 2 (next): Setting of next time point for reading date 2. Enter the new date manually or click on the description Date on the right hand side of the date entry field and use the upcoming calendar to…
-
(SW 29) Maximum Values
The recorded maximum values (power, volume flow, flow and return temperature (max.), temperature difference (max) with their respective time points (date and time) are displayed. Integration Time: The integration time for calculating the maximum values. Possible values are 6, 15, 30 and 60 minutes…
-
(SW 29) Average Values
The recorded average values (flow and return temperature, temperature difference are displayed. The integration time for calculating the average values is shown. The values shown in this card can not be changed. The integration time can be changed under maximum values. (SW 29) Leakage Acknowledge…
-
Leakage Detection (Heat): The leakage detection for the heat circuit may be switched
on or off. With the same option dialog the accuracy / sensitivity for the leakage detection may be set: Integration Time: The integration time of the leakage detection is set with this option. Possible integration times are between 1 hour and 24 hours. Stop Time Input 1: Alternatively to the leakage…
-
Alarm Duration = 5 days
In this case the alarm output is set for 30 minutes per day for a maximum of 5 days. Afterwards, the alarm output is no longer set. The alarm may also be switched off earlier by acknowledging it Alarm Hold: If the alarm output is e.g. connected to an flow interruption valve which is cutting of the…
-
(SW 29) Monthly Log
Different month values of the last 24 months are displayed. The monthly log is not loaded automatically during a standard request, instead, the user has to request it separately. Save Values: Stores the list of end of month values into a text file. Log Day: The day of the month, at which the month…
-
(SW 29) EEPROM
The available data logger memory can be split in two parts (EEPROM memory 1 and EEPROM memory 2) with two different logging time intervals. Both memories are freely programmable. The content of the EEPROM data logger memory is displayed. This meter may store up to 7040 values split up in two memory…
-
Read EEPROM Memory 1: Reads out the EEPROM data logger memory 1
Depending on the number of time points stored this process may take up to 5 minutes (2400 baud). Therefore, it is possible to limit the number of time points to read. Read EEPROM Memory 2 : Reading memory part 2; same function as EEPROM Memory 1. Table Of Contents SHARKY SW29 Energy Meter 102
-
(SW 29) EEPROM Configuration
Set Configuration : From a list of 38 possible values up to 30 values can be selected for the configuration of each EEPROM memory logger. Set allocation : The partitioning of the memory logger can be defined in percent with a scroll bar. Table Of Contents SHARKY SW29 Energy Meter 103
-
(SW 29) Errors
A list with the latest 31 events / errors is displayed. The error log is not automatically loaded during a standard request, instead, it has to be requested separately. Only available with the professional version ! Reset Error Log: The error log is deleted completely.. Save Values: Stores the…
-
(SW 29) Display 1
Here the user may select and set the sequence of the displayed values for each of the 6 possible display loops of the meter. The list on the left hand side contains all possible display values, the list on the right hand side shows the currently programmed display sequence of the display loops. By…
-
Delete All: All display values are deleted
Notes: • It is not possible to delete the energy value [00] from display loop 1. • All changes to the right hand side list are not taken unless the user has programmed them using the Set Values For Display button. The assembly of the display values in the right hand side list is not sufficient for…
-
(SW 29) Display extended
Here the user may enable or disable certain values or loops in the display of the meter (display loop 2 – 6). Usually the meter displays the display loop 1, that is by pressing the user button on the meter the user cycles through the display values of display loop 1. The setting of display loop 1 is…
-
SHARKY SW29 Energy Meter
-
(SW 29) Tariff
The meter has got two programmable tariffs which are countin energy, time or volume. The values of the two tariffs are also stored at reading date 1 and reading date 2. Clear Tariff 1: Clear values for tariff 1. Tariff 1 Limit: Set type and limit of tariff 1. For tariff type t008 and t009 an…
-
possibility to logically AND tariff 1 and tariff 2
(SW 29) Impulses Impulse Output 1 / 2: Sets the type of the impulse outputs 1 and 2. In general the meter may output an energy or volume proportional impulse or a state output which will be set if one of the following conditions is met: • Energy: Pulse weight is the lowest digit of the energy…
-
• Volume / 10
Pulse weight is the lowest digit of the volume display / 10 (i.e. Volume = 66,98 m³ ==> Pulse weight = 1 Liter) • Volume * 10: Pulse weight is the lowest digit of the volume display * 10 (i.e. Volume = 66,98 m³ ==> Pulse weight = 100 Liter) • Volume * 100: Pulse weight is the lowest digit of the…
-
The meter count on the impulse inputs may be freely set
With the declaration of the pre-counter (value between 1 and 2000) the impulse input can be divided. Table Of Contents SHARKY SW29 Energy Meter 112
-
(SW 29) Settings
Internal parameters of the device, which may be used for trouble shooting, are displayed. You have to press the protection key inside the device to alter any of the parameters on this page (with the exception of the impulse settings). After pressing the protection key inside the device the…
-
pipe system
Temperature Sensor: Indicates the type of the temperature sensor (Pt100, Pt500). Nur in der Professionell Version verfügbar ! Days with Errors: Indicates the number of days with errors (since installation of the meter or since last reset of the error log). Error State: Indicates the currently…
-
(SW 29) Calibration
Internal parameters of the device, which may be used for trouble shooting, are displayed. You have to press the protection key inside the device to alter any of the parameters on this page. After pressing the protection key inside the device the protection level is switched from normal mode to…
-
not working properly
E.g. if the device measures +1% volume flow you may set the adjustment to -1% to compensate the error. Only available with the professional version ! Start Offset Calibration: The automatic offset calibration process for the ultrasonic measuring chamber is started. The process takes about 1 minute…
-
(SW 29) Telegram
For this meter the user may assemble his own customer specific M-Bus telegram. For meters equipped with a real data radio module this customer telegram is also used as radio telegram, therefore, the user may set the values to be transmitted via real data radio. If the customer telegram is used as…
-
• All changes to the right hand side list (customer telegram) are not taken unless the user has
programmed them using the Set customer telegram button. The assembly of the values in the right hand side list is not sufficient for activating the customer telegram. • If the list on the right hand side is completely empty, the standard customer telegram is used: • [03] Energy • [04] Volume • [05]…
-
SHARKY-VMC (BR473)
Note For the configuration of this device a special L-Bus converter is necessary, which is to be connected to the pulse output of the device. The following configuration parameters must be used: At the main screen: At the settings dialog: M-Bus point to point connection using 2400 baud and L-Bus VMC…
-
The first line of entry fields indicates
Device Type and Class: SHARKY-VMC 473 qp 0.6 M-Bus Primary Address: 0 Note: The primary address of this meter type is not alterable. M-Bus Secondary Address / Device Address: 12345678 Note: If you want to change the device address you have to enable the respective entry field at the Settings dialog.…
-
A list with the latest 128 events / errors is displayed. The error log is not automatically loaded during a standard request, instead, it has to be requested separately. Save Values: Stores the displayed list of errors into a text file. Table Of Contents SHARKY-VMC (BR473) 121
-
Internal parameters of the device, which may be used for trouble shooting, are displayed. You have to connect the protection connector at the inside of the device to alter any of the parameters on this page (Protection Level: Calibration Mode). Fabrication Number: Apart from the M-Bus device ID /…
-
not working properly
E.g. if the device measures +1% volume flow you may set the adjustment to -1% to compensate the error. Only available with the professional version ! Start Offset Calibration: The automatic offset calibration process for the ultrasonic measuring chamber is started. The process takes about 1 minute…
-
SHARKY-VMC (BR473)
-
ENERGY-HEAT and RAY-HEAT Heat
Meters Current Values The first line of entry fields indicates: Device Type: RAY-HEAT M-Bus Primary Address: 4 M-Bus Secondary Address / Device Address: 81080501 Note: You may not change the device address for this device. Manufacturer Code: HYD Version Number: 29 Medium: Heat (Outlet) Programming…
-
Date+Time set: Date and time are set to the entered value. Enter the new date
manually or click on the description Date+Time Set on the right hand side of the date entry field and use the upcoming calendar to enter the date. Synchronize With PC: Date and time are synchronized with the PC clock. Note: By clicking the description on the right hand side of the date entry field…
-
The latest reading date value (Energy Reading Date 1) and, additionally, its time point (Reading
Date 1) are displayed. Reading Date 1 (next): Setting of next time point for reading date 1. Enter the new date manually or click on the description Date on the right hand side of the date entry field and use the upcoming calendar to enter the date. th February, 29 may not be set as reading date.…
-
The energy values at the beginning of each month (1 of the month at 00:00 o’clock) for the last 12
months are displayed. Table Of Contents ENERGY-HEAT and RAY-HEAT Heat Meters 128
-
Indicates if the device must be installed at the inlet or outlet (default) of a heating pipe system
Additionally, the current volume and energy are displayed with high resolution. Only available with the professional version ! Read Volume: The current volume and the current energy of the device is read in high resolution. The volume value may be e.g. used to calibrate the device. By recording the…
-
pipe system
Table Of Contents ENERGY-HEAT and RAY-HEAT Heat Meters 130
-
RAY Heat Meter
Current Values The first line of entry fields indicates: Device Type: RAY M-Bus Primary Address: 1 M-Bus Secondary Address / Device Address: 31966793 Note: If you want to change the device address you have to enable the respective entry field at the Settings dialog. Manufacturer Code: HYD Version…
-
Heat
If the device is configured as climate counter (cooling is main energy), the current heat values are shown here. If the device is configured as heat / cooling meter only, this page will be empty. The RAY heat meter is available as pure heat / cooling meter, and also as cooling meter with climate…
-
Reading Date Values
The latest reading date value (Energy Reading Date 1) and, additionally, its time point (Reading Date 1) are displayed for the main energy as well as for the heat energy, respectively. If the device is not configured as climate meter, the heat values are not shown. Reading Date 1 (next): Setting of…
-
RAY Heat Meter
-
Clear All: All recorded maximum values are cleared
Table Of Contents RAY Heat Meter 135
-
The energy values at the beginning of each month (1 of the month at 00:00 o’clock) for the last 12
months are displayed (main energy). Table Of Contents RAY Heat Meter 136
-
Monthly Log Heat
st The energy values at the beginning of each month (1 of the month at 00:00 o’clock) for the last 12 months are displayed (heat energy for devices with climate option). Table Of Contents RAY Heat Meter 137
-
This list shows the first six errors and, additionally, the latest error having appeared during the runtime of the device. Table Of Contents RAY Heat Meter 138
-
Different internal device settings are shown
Fabrication Number: Apart from the M-Bus device ID / secondary address the device has got an internal fabrication number. By default M-Bus device ID and fabrication number are equal. However, since the user is able to change the M-Bus device ID the fabrication number is displayed here once more. The…
-
compact V Heat Meter
Current Values There is only one possibility to communicate with a compact V device. User the optical communication type and choose IrDA Opto transceiver HY-Group from the general settings dialog. The first line of entry fields indicates: Device Type: compact V M-Bus Secondary Address / Device…
-
Heat / Cold
If the device is configured as combined heat/cooling meter (heating is main energy), the current cold values are shown here. If the device is configured as a pure heat or cooling meter, this page will be empty. Temperature Limit: Setting of the limit temperature, where the device is starting to…
-
Reading Date Values
The latest reading date value (Energy Reading Date) and, additionally, its time point (Reading Date) are displayed for the main energy as well as for the second energy, respectively. If the device is not configured as combined heat/cooling meter, the second values are not shown. Reading Date…
-
In case of errors, the Error State list shows all of them. The date of the last error appearance of each register is shown in the Error Date fields. Pure heat or cooling meter are just using the main error date. The compact V knows eight different error messages. Table Of Contents compact V Heat…
-
Different internal device settings and values are shown
Energy (hi-res): The current energy in high resolution (main energy register). Volume (hi-res): The current volume in high resolution (main energy register). Energy (hi-res): The current energy in high resolution (second energy register). Volume (hi-res): The current volume in high resolution…
-
FLYPPER II and SCAMPY Water Meters
Current Values The first line of entry fields indicates: Device Type: SCAMPY-FLYPPER II M-Bus Primary Address: 4 M-Bus Secondary Address / Device Address: 12345678 Note: If you want to change the device address you have to enable the respective entry field at the Settings dialog. Manufacturer Code…
-
The latest reading date value (Volume Reading Date 1) and, additionally, its time point (Reading Date 1) are displayed. Additionally, the mid (Reading Date 3) and end (Reading Date 2) of month values for the preceding month are displayed. Reading Date 1 (next): Setting of next time point for reading…
-
The volume values at the beginning of each month (1 of the month at 00:00 o’clock) for the last 12
months are displayed. Table Of Contents FLYPPER II and SCAMPY Water Meters 147
-
If the device is equipped with non-volatile EEPROM memory, it is able to store 480 volume meter
counts at user-definable time points. These meter counts may be read and displayed at this page. If the device is not equipped with EEPROM memory, this page remains blank. Read EEPROM Memory: Read the stored values from the EEPROM logger memory (max. 480 values). Only valid values are read and…
-
FLYPPER II and SCAMPY Water Meters
-
Program RAM: The basic parameters for different devices may be programmed
Table Of Contents FLYPPER II and SCAMPY Water Meters 150
-
FLYPPER III and SCAMPY II Water
Meters Current Values The first line of entry fields indicates: Device Type: SCAMPY-FLYPPER III M-Bus Primary Address: 2 M-Bus Secondary Address / Device Address: 15051971 Note: You may not change the device address for this device. Manufacturer Code: HYD Version Number: 3A Medium: Water Programming…
-
The latest reading date value (Volume Reading Date 1) and, additionally, its time point (Reading Date 1) are displayed. Reading Date 1 (next): Setting of next time point for reading date 1. Enter the new date manually or click on the description Date on the right hand side of the date entry field…
-
The volume values at the beginning of each month (1 of the month at 00:00 o’clock) for the last 12
months are displayed. Table Of Contents FLYPPER III and SCAMPY II Water Meters 153
-
Volume (hi-res): The current volume of the device is read with high resolution
Volume (forward): The current forward volume (counted in direction of the normal flow) is indicated. Volume (reverse): The current reverse volume (counted in the opposite direction of the normal flow) is indicated. Protection Level: The current protection level of the device. Normal mode is the…
-
FLYPPER IV Water Meter
Current Values The first line of entry fields indicates: Device Type: FLYPPER IV M-Bus Primary Address: 1 M-Bus Secondary Address / Device Address: 00950602 Note: If you want to change the device address you have to enable the respective entry field at the Settings dialog. Manufacturer Code: HYD…
-
The latest reading date value (Volume Reading Date 1) and, additionally, its time point (Reading Date 1) are displayed. The reading date value is always saved at 23:59. Reading Date 1 (next): Setting of next time point for reading date 1. Enter the new date manually or click on the description Date…
-
The volume values at the beginning of each month (1 of the month at 00:00 o’clock) for the last 18
months are displayed. Table Of Contents FLYPPER IV Water Meter 157
-
This list shows the errors having appeared during the runtime of the device (with their respective time stamp). Table Of Contents FLYPPER IV Water Meter 158
-
Volume (hi-res): The current volume of the device is read with high resolution
Volume (forward): The current forward volume (counted in direction of the normal flow) is indicated. Volume (reverse): The current reverse volume (counted in the opposite direction of the normal flow) is indicated. Protection Level: The current protection level of the device. Normal mode is the…
-
Pulse Valence 2: Programming of the pulse valence for pulse output 1. The device
accepts but decimal pulse valence (e.g.: 0.0001 / 0.001 / 0.01 / 0.1). Other pulse valences will be rejected by the device during programming time and the user will receive an error. Impulse Output: The device has got two impulse outputs with different programmable functions (see the list below).…
-
ENERGIE-INT 5 Heat Meter
Notes • The ENERGIE-INT 5 is only capable of communicating with 300 Baud (production date before June 2003) or with 2400 Baud (production date after June 2003). • For programming a complete parameter set, the meter to be programmed must be read once, afterwards the parameter profile must be loaded…
-
Date+Time set: Date and time are set to the entered value. Enter the new date
manually or click on the description Date+Time Set on the right hand side of the date entry field and use the upcoming calendar to enter the date. Synchronize With PC: Date and time are synchronized with the PC clock. Note: By clicking the description on the right hand side of the date entry field…
-
The latest reading date values (Energy / Volume Reading Date 1 and Energy / Volume Reading Date
2) and, additionally, their time points (Reading Date 1 and Reading Date 2) are displayed. Reading Date 1 (next): Setting of next time point for reading date 1. Enter the new date manually or click on the description Date+Time Set on the right hand side of the date entry field and use the upcoming…
-
Maximum Values: The 3 maximum value storages may either be used to store the
maximum power or the maximum flow: Table Of Contents ENERGIE-INT 5 Heat Meter 164
-
Integration Time: The integration time in minutes for calculating the maximum values
Possible values are 5, 15, 30 and 60 minutes. Table Of Contents ENERGIE-INT 5 Heat Meter 165
-
The values of energy, volume and operating time at the last day of the month at 23:59 for the last 24 months, respectively, are displayed. The monthly log must be read separately. Read Monthly Log: The monthly log is read. In contrary to most other values the user must read the monthly log manually…
-
A list with the latest 10 errors is displayed. The error log must be read separately
Read Error Log: The error log is read. In contrary to most other values the user must read the error log manually since it is not retrieved with the standard reading of the device. Table Of Contents ENERGIE-INT 5 Heat Meter 167
-
Here you are able to select the display sequence of the device. The display has got six loops (menu
1..5 and reset menu). With a short press of the button of the device you cycle through the different menus. With a long press you enter the respective menu. Set Values For Display: The display sequence shown in the list is programmed into the device. Table Of Contents ENERGIE-INT 5 Heat Meter 168
-
The device has got two tariff counters, which count if a user-definable limit is exceeded
Tariff 1 Limit: Sets the limit of tariff 1. Tariff 2 Limit: Sets the limit of tariff 2. Type Of Tariff: Sets the type of the tariff: Table Of Contents ENERGIE-INT 5 Heat Meter 169
-
Counter
Depending on the version of the device it has got two additional pulse counters (counter A and counter B), which may be configured here. Reset: The respective pulse counter input is reset to 0. Set Unit: Sets the unit for the respective pulse counter input: Table Of Contents ENERGIE-INT 5 Heat Meter…
-
Pulse Factor A / B: Sets the pulse factor for the respective pulse counter input
Decimal Digits A / B: Sets the number of decimal digits of the respective pulse counter input: Table Of Contents ENERGIE-INT 5 Heat Meter 171
-
ENERGY-INT 6 Heat Meter
Current Values The first line of entry fields indicates: Device Type and Class: ENERGY-INT 6 qp 25 M-Bus Primary Address: 0 M-Bus Secondary Address / Device Address: 87654321 Note: If you want to change the device address you have to enable the respective entry field at the Settings dialog.…
-
manually or click on the description Set Date+Time on the right hand side of the date entry field and use the upcoming calendar to enter the date. Synchronize With PC: Date and time are synchronized with the PC clock. Note: If the meter is in normal mode, it is not possible but to set the time. The…
-
Reading Dates Values 1
The values for reading date 1 are shown. LY: last year’s value T1: tariff 1 T2: tariff 2 Reading Date 1 (next): Setting of next time point for reading date 1. Enter the new date manually or click on the description Date on the right hand side of the date entry field and use the upcoming calendar to…
-
Reading Dates Values 2
The values for reading date 2 are shown. LY: last year’s value T1: tariff 1 T2: tariff 2 Reading Date 2 (next): Setting of next time point for reading date 2. Enter the new date manually or click on the description Date on the right hand side of the date entry field and use the upcoming calendar to…
-
Integration Time: The integration time in minutes for calculating the maximum values
Possible values are 6, 15, 30 and 60 minutes and 24 hours. 60 minutes is the default integration time. Table Of Contents ENERGY-INT 6 Heat Meter 176
-
Leakage
Acknowledge Alarm: An eventual alarm is reset. Leakage Detection (Heat): The leakage detection for the heat circuit may be switched on or off. With the same option dialog the accuracy / sensitivity for the leakage detection may be set: Table Of Contents ENERGY-INT 6 Heat Meter 177
-
Integration Time: The integration time of the leakage detection is set with this option
Possible integration times are between 1 hour and 24 hours. Stop Time Input 1: Alternatively to the leakage detection for the heat circuit, the device may set an alarm if there is not a certain interval of time within the alarm integration time without an impulse on impulse input 1. Usually, a…
-
Different month values of the last 24 months are displayed. The monthly log is not loaded automatically during a standard request, instead, the user has to request it separately. Save Values: Stores the list of end of month values into a text file. Log Day: The day of the month, at which the month…
-
The content of the EEPROM data logger memory is displayed. This meter may store up to 440 time
points with different values. Due to the large size of the EEPROM data logger memory it is not loaded automatically during a standard request, instead, the user has to request it separately. Please note that the refreshing of the list, depending on the number of values read, may take from several…
-
The default value for the storage interval is 24 hours, EEPROM memory 2 is not used
Read EEPROM Memory: Reads out the EEPROM data logger memory. Depending on the number of time points stored this process may take up to 5 minutes (2400 baud). Therefore, it is possible to limit the number of time points to read. Read EEPROM Memory 2 : Reads out the EEPROM data logger memory 2. If the…
-
A list with the latest 31 events / errors is displayed. The error log is not automatically loaded during a standard request, instead, it has to be requested separately. Only available with the professional version ! Reset Error Log: The error log is deleted completely.. Save Values: Stores the…
-
Display 1
Here the user may set the sequence of the displayed values in display loop 1 of the meter. The list on the left hand side contains all possible display values, the list on the right hand side shows the currently programmed display sequence of display loop 1. By selecting values (multiple selections…
-
Display 2-6
Here the user may enable or disable certain values in the display of the meter (display loop 2 – 6). Usually the meter displays the display loop 1, that is by pressing the user button on the meter the user cycles through the display values of display loop 1. The setting of display loop 1 is…
-
The meter has got two programmable tariffs for the energy value. The energy values of the two tariffs are also stored at reading date 1 and reading date 2. Clear Tariff 1: Clear values for tariff 1. Tariff 1 Limit: Set type and limit of tariff 1. For tariff type t08 and t09 an additional minimum…
-
If the user chooses the option Time Tariff the meter will not count and display the energy for the
respective tariff, but the time in hours during which the tariff condition was met. Clear Tariff 2: Clear values for tariff 2. Tariff 2 Limit: Set type and limit of tariff 2. In addition to tariff 1 tariff 2 offers the possibility to logically AND tariff 1 and tariff 2. Table Of Contents ENERGY-INT…
-
Internal parameters of the device, which may be used for trouble shooting, are displayed. You have to press the protection key inside the device to alter any of the parameters on this page (with the exception of the impulse settings). After pressing the protection key inside the device the…
-
pipe system
Temperature Sensor: Indicates the type of the temperature sensor (Pt100, Pt500). Nur in der Professionell Version verfügbar ! Errors: Indicates the number of days with errors (since installation of the meter or since last reset of the error log). For some devices with leakage detection the error day…
-
impulse input 1 and 2
Sets the impulse weight and the current impulse counts of NOTE: If the leakage detection of the heat circuit is enabled, the setting of the meter count for impulse input 1 is disabled. Impulse Output 1 / 2: Sets the type of the impulse outputs 1 and 2. In general the meter may output an energy or…
-
• Volume * 100
Pulse weight is the lowest digit of the volume display * 100 (i.e. Volume = 66,98 m³ ==> Pulse weight = 1 m³) • Energy Tariff 1: Pulse output corresponds to tariff counter 1 Pulse weight is the lowest digit of the energy display • Energy Tariff 2: Pulse output corresponds to tariff counter 2 Pulse…
-
Internal parameters of the device, which may be used for trouble shooting, are displayed. You have to press the protection key inside the device to alter any of the parameters on this page. After pressing the protection key inside the device the protection level is switched from normal mode to…
-
not working properly
E.g. if the device measures +1% volume flow you may set the adjustment to -1% to compensate the error. Only available with the professional version! Valence Impulse Output: The pulse weight of the output pulses which are available by using the pulse output module without galvanic isolation (maximum…
-
Telegram
For this meter the user may assemble his own customer specific M-Bus telegram. For meters equipped with a real data radio module this customer telegram is also used as radio telegram, therefore, the user may set the values to be transmitted via real data radio. If the customer telegram is used as…
-
• All changes to the right hand side list (customer telegram) are not taken unless the user has
programmed them using the Set customer telegram button. The assembly of the values in the right hand side list is not sufficient for activating the customer telegram. • If the list on the right hand side is completely empty, the standard customer telegram is used: • [03] Energy • [04] Volume • [05]…
-
SCYLAR INT7 Energy Meter
(INT7) Current Values The first line of entry fields indicates: Device Type and Class: Scylar Int7 M-Bus Primary Address: 0 M-Bus Secondary Address / Device Address: 12345678 Note: If you want to change the device address you have to enable the respective entry field at the Settings dialog.…
-
manually or click on the description Set Date+Time on the right hand side of the date entry field and use the upcoming calendar to enter the date. Synchronize With PC: Date and time are synchronized with the PC clock. Note: By clicking the description on the right hand side of the date entry field…
-
(INT7) Reading Dates Values 1
The values for reading date 1 are shown. LY: last year’s value T1: tariff 1 T2: tariff 2 Reading Date 1 (next): Setting of next time point for reading date 1. Enter the new date manually or click on the description Date on the right hand side of the date entry field and use the upcoming calendar to…
-
(INT7) Reading Dates Values 2
The values for reading date 2 are shown. LY: last year’s value T1: tariff 1 T2: tariff 2 Reading Date 2 (next): Setting of next time point for reading date 2. Enter the new date manually or click on the description Date on the right hand side of the date entry field and use the upcoming calendar to…
-
(INT7) Maximum Values
The recorded maximum values (power, volume flow, flow and return temperature (max.), temperature difference (max) with their respective time points (date and time) are displayed. Integration Time: The integration time for calculating the maximum values. Possible values are 6, 15, 30 and 60 minutes…
-
(INT7) Average Values
The recorded average values (flow and return temperature, temperature difference) are displayed. The integration time for calculating the average values is shown. The values shown in this card can not be changed. The integration time can be changed under maximum values. (INT7) Leakage Acknowledge…
-
Leakage Detection (Heat): The leakage detection for the heat circuit may be switched
on or off. With the same option dialog the accuracy / sensitivity for the leakage detection may be set: Integration Time: The integration time of the leakage detection is set with this option. Possible integration times are between 1 hour and 24 hours. Stop Time Input 1: Alternatively to the leakage…
-
Alarm Duration = 5 days
In this case the alarm output is set for 30 minutes per day for a maximum of 5 days. Afterwards, the alarm output is no longer set. The alarm may also be switched off earlier by acknowledging it Alarm Hold: If the alarm output is e.g. connected to a flow interruption valve, which is cutting off the…
-
(INT7) Monthly Log
Different month values of the last 24 months are displayed. The monthly log is not loaded automatically during a standard request, instead the user has to request it separately. Save Values: Stores the list of end of month values into a text file. Log Day: The day of the month, at which the month…
-
(INT7) EEPROM
The available data logger memory can be split in two parts (EEPROM memory 1 and EEPROM memory 2) with two different logging time intervals. Both memories are freely programmable. The content of the EEPROM data logger memory is displayed. This meter may store up to 7040 values split up in two memory…
-
Read EEPROM Memory 1: Reads out the EEPROM data logger memory 1
Depending on the number of time points stored this process may take up to 5 minutes (2400 baud). Therefore, it is possible to limit the number of time points to read. Read EEPROM Memory 2 : Reading memory part 2; same function as EEPROM Memory 1. Table Of Contents SCYLAR INT7 Energy Meter 205
-
(INT7) EEPROM Configuration
Set Configuration : From a list of 38 possible values up to 30 values can be selected for the configuration of each EEPROM memory logger. Set allocation : The partitioning of the memory logger can be defined in percent with a scroll bar. Table Of Contents SCYLAR INT7 Energy Meter 206
-
(INT7) Errors
A list with the latest 31 events / errors is displayed. The error log is not automatically loaded during a standard request, instead, it has to be requested separately. Only available with the professional version! Reset Error Log: The error log is deleted completely.. Save Values: Stores the…
-
(INT7) Display 1
Here the user may select and set the sequence of the displayed values for each of the 6 possible display loops of the meter. The list on the left hand side contains all possible display values, the list on the right hand side shows the currently programmed display sequence of the display loops. By…
-
Delete All: All display values are deleted
Notes: • It is not possible to delete the energy value [00] from display loop 1. • All changes to the right hand side list are not taken unless the user has programmed them using the Set Values For Display button. The assembly of the display values in the right hand side list is not sufficient for…
-
(INT7) Display extended
Here the user may enable or disable certain values or loops in the display of the meter (display loop 2 – 6). Usually the meter displays the display loop1, that is by pressing the user button on the meter the user cycles through the display values of display loop 1. The setting of display loop 1 is…
-
SCYLAR INT7 Energy Meter
-
(INT7) Tariff
The meter has got two programmable tariffs, which are counting energy, time or volume. The values of the two tariffs are also stored at reading date 1 and reading date 2. Clear Tariff 1: Clear values for tariff 1. Tariff 1 Limit: Set type and limit of tariff 1. For tariff type t008 and t009 an…
-
possibility to logically AND tariff 1 and tariff 2
(INT7) Impulses Impulse Output 1 / 2: Sets the type of the impulse outputs 1 and 2. In general the meter may output an energy or volume proportional impulse or a state output which will be set if one of the following conditions is met: • Energy: Pulse weight is the lowest digit of the energy display…
-
• Volume / 10
Pulse weight is the lowest digit of the volume display / 10 (i.e. Volume = 66,98 m³ ==> Pulse weight = 1 Liter) • Volume * 10: Pulse weight is the lowest digit of the volume display * 10 (i.e. Volume = 66,98 m³ ==> Pulse weight = 100 Liter) • Volume * 100: Pulse weight is the lowest digit of the…
-
The meter count on the impulse inputs may be freely set
With the declaration of the pre-counter (value between 1 and 2000) the impulse input can be divided. pulse value of the flow sensor The pulse value of the pulses coming from the connected flow sensor can be defined. ATTENTION: If you change this value, you have to recalibrate the meter and redesign…
-
(INT7) Settings
Internal parameters of the device, which may be used for trouble shooting, are displayed. You have to press the protection key inside the device to alter any of the parameters on this page (with the exception of the impulse settings). After pressing the protection key inside the device the…
-
pipe system
Temperature Sensor: Indicates the type of the temperature sensor (Pt100, Pt500). Only available with the professional version! Days with Errors: Indicates the number of days with errors (since installation of the meter or since last reset of the error log). Error State: Indicates the currently…
-
(INT7) Calibration
Internal parameters of the device, which may be used for trouble shooting, are displayed. You have to press the protection key inside the device to alter any of the parameters on this page. After pressing the protection key inside the device the protection level is switched from normal mode to…
-
not working properly
E.g. if the device measures +1% volume flow you may set the adjustment to -1% to compensate the error. Table Of Contents SCYLAR INT7 Energy Meter 219
-
(INT7) Telegram
For this meter the user may assemble his own customer specific M-Bus telegram. For meters equipped with a real data radio module this customer telegram is also used as radio telegram, therefore, the user may set the values to be transmitted via real data radio. If the customer telegram is used as…
-
• All changes to the right hand side list (customer telegram) are not taken unless the user has
programmed them using the Set customer telegram button. The assembly of the values in the right hand side list is not sufficient for activating the customer telegram. • If the list on the right hand side is completely empty, the standard customer telegram is used: • [03] Energy • [04] Volume • [05]…
-
HYDRUS water meter
The Hydrus water meter is available in three variants: radio, M-Bus and pulse. The HydroSet display may vary according to the variant used. (HYDRUS) Current Values The following parameters are shown in the top line of the display: Meter type: Hydrus with Q3 and interfaces M-Bus primary address: 124…
-
The current values in the meter are shown below
Set Date+Time: For setting the date and time to the given value. Type in the date or click Set Date+Time to display the calendar input window. Synchronize With PC: For adjusting the date and time to the current time of the computer. Note: A mouse click on the lettering after the date input field…
-
(HYDRUS) Reading Date Values
The values for reading date 1 are shown. pY: previous year Reading Date 1 (next): For setting the next reading date for reading date 1. Type in the reading date or click Date to display the calendar input window. Notes: The reading date values always refer to the end of the day, i.e. to 23.59 hours…
-
(HYDRUS) Log Storage
This window shows the contents of the log storage memory. The log storage is not read automatically in a standard request, but must be read explicitly using the relevant button. The meter can store maximum 32 data records. These comprise: - Maximum flow - Minimum flow - Total volume - Return volume…
-
(HYDRUS) Error
A list of the last 31 events / errors is shown. The error log is not read automatically in a standard request, but must be read explicitly using the relevant button. Only available in the professional version! Delete Error Log: The error log is deleted completely. Save Values: Saves the errors shown…
-
(HYDRUS) Display
This tab is for programming the selection of fields and the sequence of displayed values for the meter display. The left list contains the possible values for the display and the list on the right shows the previously defined display values. The display configuration is set by selecting values…
-
(HYDRUS) Settings
Shows internal parameters of the meter that can be used for tasks such as fault diagnosis. To change parameters on this tab, the calibration button inside the meter must always be pressed. Pressing the calibration button in the meter changes the programming protection level from normal mode to…
-
Total volume: The total volume in conjunction with the direction information (pulse output 2) represents in principle the total of the forward and return flows at any time. No return pulses are buffered and compensated with a subsequent forward flow. If pulse output 2 for the direction indication is…
-
(HYDRUS) Calibration
This tab shows other internal parameters of the meter, which can be used for tasks such as fault diagnosis. To change parameters on this tab, the calibration button inside the meter must always be pressed. Pressing the calibration button in the meter changes the programming protection level from…
-
Offset Calibration On
(HYDRUS) Telegram The meter offers the user the possibility of compiling a customer-specific M-Bus telegram of his choice. This customer-specific M-Bus telegram is also used as a radio telegram if the meter is used with a Real Data Radio Module, i.e. the values to be sent over Real Data Radio are…
-
• Each value can only be used once in the customer telegram
• Changes to the customer telegram (right list) are not transferred to the meter until the Set Customer Telegram button has been pressed. Compiling the values in the right list is not sufficient to activate the customer telegram. A different telegram can be set as standard if requested by the…
-
How Do I …?
Communication Using Opto Transceiver 1. Connect the opto transceiver to a free serial port of your PC. 2. Put the opto transceiver onto the device to read out. 3. Select "Optical Communication (According To Settings)". 4. At the Settings dialog select “ZVEI Optotransceiver”. 5. Select the correct…
-
Communication M-Bus (Only One Device)
1. Connect the M-Bus repeater (e.g. HYDRO-CENTER®) to a free serial port of your computer. 2. Connect the M-Bus output of the repeater to the M-Bus connector of the device to read out. Please verify that only one device is connected to the M-Bus repeater. 3. Select "Serial Cable (directly)". 4.…
-
Communication M-Bus (Primary Address)
1. Connect the M-Bus repeater (e.g. HYDRO-CENTER®) to a free serial port of your computer. 2. Connect the M-Bus output of the repeater to the M-Bus connector of the device to read out. 3. Select "Serial Cable (directly)". 4. Select the correct serial port (e.g. COM2). 5. Select 2400 Baud. 6. Select…
-
Set New Primary Address
1. Establish a connection to the device to read out (as described in Communication Using Opto Transceiver, Communication M-Bus (Only One Device), Communication M-Bus (Secondary Address), Communication M-Bus (Primary Address), or Communication Using A Dial-Up Phone Line). 2. Enter a new primary…
-
Trouble Shooting
No communication with the meter If there is no communication between the computer and the meter, please check the following points: • the correct type of the communication (optical, M-Bus, modem) • serial port (COM1, COM2, …) • communication speed (usually 300 or 2400 baud) • selection type for…
-
Afterwards it is mandatory to restart Windows, even though there might not be a message
Table Of Contents Trouble Shooting 238