NuAire Ultra Low Temperature Freezer
-
2 Product Characteristics and Model Meaning
2.1. Main Functions and Characteristics of the Product 2.1.1. The applicable temperature scope in the ultra-low temperature freezer is -40~-86℃; 2.1.2. The freezer has the balanced internal and external pressure design and its doors can be opened and closed easily; 2.1.3. The display panel is…
-
3. Product Promotional Points
3.1 Temperature Control: The temperature is controlled by computer and displayed digitally, its adjustment unit is 1℃ and the adaptable temperature scope is -40~-86℃; 3.2 Safety Control: 3.2.1 Multiple failure alarm (high/low temperature alarm, sensor alarm, door opening alarm, hot condenser alarm…
-
4. Product Appearance Structure & Internal Structure Diagram
4.1 Energy saving series: NU-99486JE NU-99578JE/578JGA/578JGA NU-99728JE/728JGA/728JGA Front View 4.2 Internal Structural View Rear View Slim line series:NU-99338JE/338JG/338JGA Front View 4.3 NU-99828JE/828JG Internal Structural View Chest series : NU-99420JE Front View OM0290 Revision 3 Rear View…
-
5. Product Component Structure & Names
5.1 Display: NU-99486JE NU-99578JE/578JGA/578JGA NU-99420JE NU-99728JE/728JGA/728JGA NU-99420JE NU-99828JE/828JG NU-99420JG NU-99338JE/338JG/338JGA NU-99420JG High temp. alarm Low temp. alarm Silence alarm Probe failure Locked Low battery Network Hot condenser Run Regulator High ambient temp. Backup…
-
OM0290 Cabinet Structure External door:PU foam with VIP Inner door:PU foamed External door seals can be replaced. Inner door latch:remove the side screws when replaced Inner door seals can be replaced. One piece cabinet frame Recorder mounting panel Air intake grill Main cabinet sensor shield 5V…
-
External door
handle Foam internal door Internal door sealing strip External door sealing strip Internal door handle OM0290 Revision 3 OM0290 Balance hole Magnetic control switch box for door opening alarm Threading hole at the back of the freezer, four-channel structure Page 8 of 60 Release Date: 4/13/2022
-
NU-99728JE/728JGA/728JGA NU-99828JE/828JG NU-99338JE/338JG/338JGA Remote alarm terminal Control board RS-232/485Port Location (Not Shown) Condenser fan Battery charge switch Power switch High temp stage Relay Battery Low temp stage Relay Terminal strip Filter Condenser fan OM0290 Revision 3 Page 9…
-
Board check button: press this button to
observe if all relay lights will take turn to light up. Connector for display board OM0290 USB signal connector Temperature adjustment screw Voltage calibration screw Door alarm connector Main sensor connector Heat exchanger connector Battery charge connector (white) Condenser sensor connector…
-
Junction box: NU-99420JE NU-99420JG Battery Control board USB power board Power switch OM0290 Revision 3 OM0290 Compressor relay Terminal strip Battery switch Page 11 of 60 Remote alarm contacts Filter RS-232/485Port Location (Not Shown) Release Date: 4/13/2022
-
OM0290 Refrigeration 5.4.1 Refrigeration System :NU-99486JE NU-99578JE/578JGA/578JGA NU-99728JE/728JGA/728JGA NU-99828JE/828JG High stage liquid accumulator NU-99338JE/338JG/338JGA Perimeter halo tube High Temp stage condenser High temperature filter dryer Low temp stage oil separator Low…
-
Refrigeration system: NU-99420JE OM0290 NU-99420JG High temperature compressor High stage liquid accumulator High stage filter dryer Condenser Condenser Low temperature compressor Low temperature stage filter dryer OM0290 Revision 3 Oil separator Page 13 of 60 Release Date: 4/13/2022
-
6. Product Use Introduction & Product Improvement Introduction
6.1 Unlocking: Press the key “UP” or “DOWN”. When “00” is displayed on the display panel, adjust the key "UP" to “06” (“06” is the default password) and then press the key “ SET” for 3 seconds. When the lock indicator turns off, the unit is unlocked. 6.2 Temperature Setting Operation: Pressing the…
-
6.4 Failure Code Inquiry
1) When the machine detects that the result of the condenser temperature minus the ambient temperature is ≥13℃ (for 5min. continuously), the “hot condenser” indicator turns on and the machine makes the relevant silence alarm. After the power supply is on, the indicator does not turn on and the…
-
The heat exchanger
probe foams at the bottom inside the freezer. E3 at the terminal of the heat exchanger sensor 7) When the temperature detected by the heat exchanger sensor is ≥90℃, the alarm displays E4. Remark: (If the product has no heat exchanger sensor from -25℃ to -50℃, failures E3 and E4 will not happen to…
-
Tempera
-
7.Main Technical Parameters of the Product
-
Model
NU-99828JE OM0290 NU-99828JG Technical Data Cabinet Type Climate Class Controller Display Temp Sensor Parameters Capacity(L) Power Supply(V/Hz) 220~240V/50Hz 208~230V/60Hz Power(W) 1650 1650 Rated Current 10 10 Stable Current 7.5 7.5 Single door, upright N Microprocessor control LED Display with 1 C…
-
Model
NU-99420JE OM0290 NU-99420JG Technical Data Cabinet Type Single door, chest Climate Class N Controller Microprocessor control Display LED Display with 1 C increment Temp Sensor PT100 Parameters Capacity(L) 420 Power Supply(V/Hz) 220~240V/50Hz 208~230V/60Hz Power(W) 1540 1540 Rated Current 7.5 7.5…
-
7.2 Product Nameplate
For example NU-99338JG OM0290 Revision 3 Page 21 of 60 Release Date: 4/13/2022
-
8. Case Prevention Measures, Product Use and Daily Maintenance
Knowledge 8.1 Refrigeration Principle of ULT Freezer: The ultra-low temperature freezer is equipped with a cascade refrigerating system normally composing of two or more refrigerating systems, i.e. a low temperature stage and high temperature stage. For the high temperature stage, the…
-
8.5 Refrigerant Charging Process
8.5.1 Preparation 1) Refrigerant (R290 and R170.) 2) Hansen valve 3) Standard electronic scale (error ≤±2g) 4) Vacuum pump (≥4L) 5) Pressure gauge (high and low pressures shall be read at the same time) See Fig. 1. 8.5.2 System Cutoff 1) Cut off the failure system from the end of the process pipe…
-
8.5.4 Charging Refrigerant
1) High temperature level perfusion, machine should be in stop state, reperfusion, refrigerant R290 perfusion measure implemented according to the attached form! 2) When the low temperature refrigerant is filled, the machine shall also be off. Firstly, fill R290 refrigerant and then fill R170…
-
ULT Full Series Refrigerant Charge
-
9.1 Circuit Diagram
OM0290 Revision 3 Page 26 of 60 Release Date: 4/13/2022
-
10. Specific Control, Work Principle and Parameters
Nuaire ULT’s have the temperature control system with a LED computer panel for controlling the switch-on/off state of the compressor. 10.1 Temperature control system with a LED computer panel for controlling the switch-on/off state of the compressor 1) In order to regulate the set value, firstly…
-
A. “High temperature indicator: if the indicator turns on, the alarm will be displayed because the
temperature in the freezer is higher than the maximum temperature set value; B. “Low temperature indicator: if the indicator turns on, the alarm will be displayed because the temperature in the freezer is lower than the minimum temperature set value; C. "Door open" indicator light: light flash said…
-
10.4 Battery Control Requirement
There is a battery power switch on the equipment, and the switch shall be opened before normal startup. As soon as the AC power required by the equipment is switched on, the equipment will enter the normal working state. When the AC power is normally supplied, the equipment can be charged to the…
-
10.8 Compressor Control
When power is just switched on, the time will delay according to the setting of the user (the default delay is 1min.). If the freezer meets the switch-on condition, the high temperature compressor will be started. After that, if the heat exchanger sensor temperature is not higher than -20℃ in 10…
-
button or "▽" button to display circularly from 01 to 12, after selecting the current month, press the "SET button to save. Thereafter, the inside temperature display area steadily shows "3P"; press "SET" button and the inside temperature display area shows Date with flashing (factory default 01)…
-
11. Refrigeration Cycle Prospective and Plan
11.1 Refrigeration Principle The ultra-low temperature freezer is equipped with a cascade refrigerating system normally composing of two or more refrigerating systems, i.e. low temperature stage and high temperature stage. For the high temperature stage, the medium-temperature refrigerant is used…
-
12. Installation & Disassembly Process
12.1 Installation Position 12.1.1. Prevented from direct sunshine 12.1.2. Good surrounding air ventilation condition 12.1.3. Prevented from a lot of dust 12.1.4. Prevented from mechanical vibration or shock 12.1.5. Ambient temperature: 5℃~28℃; maximum ≦ 32℃; the most ideal temperature: 18℃~25℃; the…
-
A
Fastener A Fastener B Fastener A D C B 12.3 The Product Enters the Placement Room. 12.3.1 Difficulty in entering the door: If it is impossible to enter the room without the pallet because of the door height in the room where the user wants to place the refrigerator, the user can consider…
-
③Open the external door by 180º
After the product enters the door, assemble the front cover of the compartment to its original condition. Note: The front cover shall be reassembled in place. Otherwise, safety and performance problems might be caused! When the freezer is carried, its protection box shall neither be collided nor…
-
Our machine refrigerates from top to bottom, so the temperature at the top is lower than the temperature at the bottom, the difference between the upper and lower temperatures is large, the temperature difference of similar international products is 5~8ºC, and the temperature difference of our…
-
13.2 Common Q&A
Question 1. The high temperature compressor cannot start. 2. The low temperature compressor cannot start. 3. Fan failures OM0290 Revision 3 Reason Analysis Maintenance Measures 1. Power switch or fuse failure Use the multi-meter to measure the switch or fuse resistance and check if the switch is…
-
4. High
temperature in the compartment OM0290 2. The fan blade is blocked by foreign matters. Check if any fan blade collides the wall, makes noise and not equipped when it rotates, troubleshoot the blade and remove surrounding foreign matters and keep the blades running normally. 3. Fan damage Use the…
-
9. Hot
condenser alarm When the result of the condenser temperature detected by the condenser probe is ≥13 ℃ (for 5min. constantly), an alarm happens. OM0290 1. Open the front grille and clean the filter screen. 2. Check if the condenser probe is too close to the outlet of the condenser. 3. Use the…
-
High Temperature Alarm Solution
Freezer display temp ≥set temp. for high temp. alarm (set temp. for high temp. alarm≥set temp. for running +5℃) No (normal) Yes Check the control circuit and LED lamps on the display panel. The “Alarm” indicator flickers; The “High-temp. Alarm” indicator flickers; Yes No . The freezer makes an…
-
Low Temperature Alarm Solution
Freezer display temp.≤ set temp. for low temp. alarm (set temp. for low temp. alarm ≤ set temp. for running -5℃) No (normal) Yes The “Alarm” indicator flickers; The “Low-temp. Alarm” indicator flickers; Check the control circuit and LED lamps on the display panel. Yes No The freezer makes an…
-
Alarm code "E0" is not
relieved; continue checking Code "E0" alarm is canceled; the fault is eliminated. Ambient temperature sensor is poorly connected to the display panel. Plug in the ambient temperature sensor again. Alarm code "E0" is not relieved; continue checking "E0" code alarm is canceled; the fault is…
-
Alarm code "E1" is not
relieved; continue checking Code "E1" alarm is canceled; the fault is eliminated. Condenser sensor is poorly connected to the control panel; plug in the condenser sensor again. re sens or Alarm code "E1" is not relieved; continue checking Code "E1" alarm is canceled; the fault is eliminated.…
-
Alarm code "E2" is not
relieved; continue checking Code "E2" alarm is canceled; the fault is eliminated. Inside-box sensor P T100 is poorly connected to the control panel; plug in and secure the sensor PT100 again. Alarm code "E2" Alarm code "E2" is not relieved; continue checking Code "E2" alarm is canceled; the fault is…
-
Alarm code "E3" is not
relieved; continue checking Code "E3" alarm is canceled; the fault is eliminated. Heat exchange sensor is poorly connected to the control panel; plug in the heat exchange sensor again. Alarm code "E3" is not relieved; continue checking Code "E3" alarm is canceled; the fault is eliminated. Heat ex…
-
Solution to "Fans fail to work"
OM0290 Revision 3 Page 46 of 60 Release Date: 4/13/2022
-
Solution for checking high-temperature compressor failing to start up
OM0290 Revision 3 Page 47 of 60 Release Date: 4/13/2022
-
Solution for checking low-temperature compressor failing to start up
OM0290 Revision 3 Page 48 of 60 Release Date: 4/13/2022
-
Solution to condenser alarm fault
OM0290 Revision 3 Page 49 of 60 Release Date: 4/13/2022
-
Solution to display panel NG fault
OM0290 Revision 3 Page 50 of 60 Release Date: 4/13/2022
-
14. Post-maintenance normal cooling curve data of refrigeration system