Service and Maintenance Manual i.Series® i.Series iUF118 iUF126 i.Series iUF116 iUF124 360173-A/C
-
Document History
Revision Date CO Supersession Revision Description A 29 DEC 2014* 10184 n/a Initial release. B 05 JAN 2016 11402 B supersedes A Added instructions for removal of compressor restraints. Updated initial charge values in Section 10.1. C 26 FEB 2017 12665 C supersedes B Reformatted content for…
-
Contents
About this Manual. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 1.1 Safety Precautions and Symbols . . . . . . . . . . . . . . . . . . . . . . . . . .…
-
Service and Maintenance Manual About this Manual This manual provides information on how to use the i.Series® Ultra-Low Freezer. It is intended for use by end users of the freezer and authorized service technicians. Models are indicated by a distinguishing model number that corresponds to the…
-
Avoiding Injury
Review safety instructions before installing, using, or maintaining the equipment. ♦♦ Before moving unit, remove contents from the chamber. ♦♦ Before moving unit, ensure door is closed and latched, and casters are unlocked and free of debris. ♦♦ Before moving unit, disconnect the AC power cord and…
-
Installation and Configuration 2.1 Location Requirements Service and Maintenance Manual ♦♦ Has a dedicated 15 A grounded circuit with dedicated single point receptacle meeting the electrical requirements listed on the product specification label. ♦♦ Is clear of direct sunlight, high temperature…
-
Service and Maintenance Manual Connect Back-up Power The monitoring system has a back-up battery system, enabling a period of continuous monitoring if power is lost. Battery life varies by manufacturer as well as voltage level remaining. Providing full power is available and no battery-related…
-
Additional Sensors
Additional temperature sensors may be installed through one of the ports on the rear wall of the cabinet. The ports are located in the top-left corner and bottom-left corner. The ports consist of two plastic caps - one inside the cabinet and one on the outside. The cabinet insulation fills the void…
-
Install / Replace Chart Paper
Notes • For accurate temperature reading, ensure the current time is aligned with the time line groove when chart knob is tightened • Contact Helmer Customer Service to reorder chart paper; part number 800197-1 (52 sheets). Chart recorder stylus and time line groove 1. Press and hold C button. When…
-
Service and Maintenance Manual Back-up Refrigeration Systems (optional) A back-up refrigeration system may be installed to ensure the chamber temperature remains below a critical level in the event of an extended power failure or an equipment failure. If the chamber temperature rises to a…
-
Service and Maintenance Manual Configure Storage CAUTION • Use appropriate gloves when handling cold interior components and stored inventory. • Keep hands away from pinch points when closing the door. • Before moving shelves, ensure they are empty for safe lifting. • Maximum shelf load is not to…
-
Service and Maintenance Manual Controls i.Series® Ultra-Low freezers are equipped with the i.C³ monitoring and control system. The i.C³ system combines temperature monitoring and control into a single user interface. 3.1 Home Screen and Screensaver The Home Screen is the default screen and is…
-
Service and Maintenance Manual Alarm Conditions If an alarm condition is met, an alarm activates. Some alarms are visual only; others are visual and audible. Some alarms are sent through the remote alarm interface. The table indicates if an alarm is audible (A), visual (V), or sent through the…
-
Service and Maintenance Manual Settings > Through the i.C³ monitoring and control system, current settings may be viewed and changed. To view settings, touch i.C³ APPS, Settings. Use a touch-drag motion to scroll up or down to display additional settings. Settings screens Note • If the Settings…
-
Device Control Settings
Device control settings are programmed at the factory. Setpoints can be viewed and changed through the i.C³ monitoring and control system. To view the temperature setpoint, touch i.C³ APPS, Settings, Device Control Settings. Device Control Settings screen Table 3. Setpoints Setting Initial Factory…
-
Alarm Settings screens
Change an Alarm Setting 1. Touch i.C3 APPS, Settings. 2. Enter the Settings password (default password is “1234”). 3. Scroll down and touch Alarm Settings. 4. Touch the minus (-) or plus (+) on the corresponding Setpoint spin box to change an alarm setpoint. 5. Touch the minus (-) or plus (+) on the…
-
Service and Maintenance Manual Sensor Calibration > Sensor calibration values are programmed at the factory. Calibration values can be viewed and changed through the i.C³ monitoring and control system. To view calibration settings, touch i.C³ APPS, Settings and scroll down to Sensor Calibration.…
-
Calibrate control/primary monitor probe
1. Move contents within the freezer to gain access to the probe located in the bottom compartment against the rear wall. 2. Using a zip-tie, attach the independent thermometer to the cover surrounding the probe. 3. Close the door and allow the chamber temperature to stabilize for a minimum of five…
-
Factory Default Settings
Settings listed below may be simultaneously returned to factory default values. Note The factory default settings may not be the same as the settings that were factory-calibrated before the freezer was shipped. Table 6. Restored Settings Setting Restored Value Home Screen Application Icons i.C³…
-
Restore Settings
1. Touch the i.C3 Settings icon. 2. Scroll down and touch the Restore Factory Settings button. The Restore Factory Settings confirmation box appears. 3. Touch to confirm, or to cancel. Edit Factory Settings Several of the i.C3 screens and functions are configured at the factory. The screens and…
-
Service and Maintenance Manual Maintenance Maintenance tasks should be completed according to the schedule below. CAUTION • Maintenance should only be performed by trained refrigeration technicians. • Review all safety instructions prior to performing maintenance. Notes • It is important to ensure…
-
Service and Maintenance Manual Alarm Tests > Test alarms to ensure they are working correctly. The freezer has alarms for primary monitor high temperature, primary monitor low temperature, door open (time), high ambient temperature, low ambient temperature, and power failure. CAUTION • Use…
-
Power Failure Alarm Test
During a power failure, the power failure alarm activates and the back-up battery provides power to the monitoring system. Note Do not switch the back-up battery switch Off during the power failure test. Test the power failure alarm 1. Touch, i.C³ APPS, Settings, Alarm Settings. 2. Touch the minus…
-
Low Ambient Temperature Alarm Test
Test the Low Ambient Temperature Alarm 1. Touch, i.C³ APPS, Settings, Alarm Settings 2. Scroll down and record the low ambient alarm setpoint. 3. Touch the minus (-) on the Low Ambient Time Delay setting spin box to change the setting to 0 minutes. 4. Lift the condenser grill from the magnetic…
-
Replace Monitoring System Back-up Battery
The monitoring system back-up battery is a serviceable part and available by contacting Helmer Technical Service. The battery is located behind the electrical compartment access panel below the chamber. Refer to the installation instruction included with the battery service kit. WARNING Removal of…
-
Defrost and Clean Chamber
WARNING Do not use a secondary heat source to defrost the freezer. The use of an electrical heat source (such as a heat gun) could create an electrocution hazard if the user comes into contact with water from the defrosted chamber. CAUTION • Use appropriate gloves when handling cold interior…
-
Service and Maintenance Manual Clean Ultra-Low Freezer WARNING Disconnect the freezer from AC power prior to cleaning the condenser with a liquid cleaner to prevent an electrocution hazard. CAUTION • Avoid sharp edges when removing and installing the condenser grill and filter. • Use appropriate…
-
Service 5.1 Refrigerant Service and Maintenance Manual CAUTION • Review all safety instructions prior to recharging refrigerant. • Maintenance should only be performed by trained refrigeration technicians. • Avoid sharp edges when working inside the refrigeration compartment. NOTICE The…
-
Service and Maintenance Manual Access Control Solenoid In the event of an Access Control Solenoid failure, the freezer must be accessed with the override key. The Access Control Solenoid is a serviceable assembly installed in the door. Replace Access Control Solenoid Contact Helmer Technical Service…
-
Align door flush with cabinet
1. Using a #2 Phillips screwdriver, remove the screws securing the hinge covers. Remove hinge covers and set screws and hinge covers aside. 2. Hold a straight-edge (ruler) on the side of the freezer, spanning the door and cabinet, adjacent to the hinge being adjusted. 3. Using a #20 Torx driver…
-
Service and Maintenance Manual Troubleshooting CAUTION • Review all safety instructions prior to troubleshooting. • Troubleshooting should only be performed by trained refrigeration technicians. NOTE To order Replacement Parts, contact Helmer Technical Service (1-800-743-5637). 6.1 General Operation…
-
Problem
Possible Cause Action The chart recorder is not marking the temperature. The chart paper knob is not tight. Tighten the chart paper knob. Stylus pressure is not correct. Confirm stylus is pressing firmly against paper. The chart recorder motor has stopped running. Remove the chart paper and press…
-
Problem
Possible Cause Action The chamber temperature does not reach or stabilize at setpoint. Ambient air temperature is too high. Confirm freezer location meets requirements. Warm product was placed at sensor. Move product and monitor chamber temperature. Wait to see if temperature has stabilized or…
-
Problem
Service and Maintenance Manual Possible Cause Action Both compressors running. Check cascade temperature - touch i.C3 settings (user password) / device status and history - Cascade temperature maintaining below -20ºC BUT greater than -50ºC. Normal. Allow to stabilize. Cascade temperature -50ºC or…
-
Problem
Possible Cause Action The Condenser Temperature alarm is active. Condenser filter is dirty. Check the condenser filter. Clean if necessary. Side panels missing from cabinet base. Reinstall side panels to ensure proper airflow across condenser and compressors. Condenser fan is not running. Check the…
-
Parts 7.1 Exterior Front and Door Service and Maintenance Manual Notes • Before replacing parts, protect items in freezer from extended exposure to adverse temperature. • Allow freezer temperature to stabilize at setpoint after replacing parts or after extended door opening. • The i.C³ display…
-
Service and Maintenance Manual Interior Cabinet NOTICE • When removing or replacing storage racks, do not allow the storage rack to set on the top edge of a partially-open interior door. • To avoid damage to the interior door hinges, do not apply upward or downward force to the interior doors. A D B…
-
Service and Maintenance Manual Electrical Panel CAUTION • Disconnect the freezer from AC power before removing the access panel. • Avoid sharp edges when working inside the electrical compartment. Note The i.C³ control board is sensitive to static electricity and can be damaged by electrostatic…
-
Service and Maintenance Manual Refrigeration Components and Exterior Rear CAUTION • Disconnect the freezer from AC power before removing the access panel. • Avoid sharp edges when working inside the refrigeration compartment. Note The TXV (thermal expansion valve) is not a serviceable component. To…
-
Schematics 8.1 Electrical Schematic 360173-A/C Service and Maintenance Manual 41
-
Service and Maintenance Manual Refrigeration Schematic AMBIENT AIR INTAKE RTD2 AIR FILTER high pressure gas (hot) HIGH STAGE COMPRESSOR low pressure gas (warm) TC2 high pressure gas (warm / desuperheated) high pressure gas R-404A RTD1 CONDENSER FAN HIGH STAGE RTD5 CONDENSER SUCTION LINE HEAT…
-
Appendix A: Compliance
Regulatory Compliance This product is certified to applicable UL and CSA standards by a NRTL. WEEE Compliance The WEEE (waste electrical and electronic equipment) symbol (right) indicates compliance with European Union Directive WEEE 2002/96/EC and applicable provisions. The directive sets…
-
Appendix B: Warranty
Rel.i™ Product Warranty USA and Canada For technical service needs, please contact Helmer at 800-743-5637 or www.helmerinc.com. Have the model and serial number available when calling. Rapid Resolution When a warranty issue arises it is our desire to respond quickly and appropriately. The service…
-
360173-A/C
Service and Maintenance Manual 45
-
Helmer Scientific
14400 Bergen Boulevard, Noblesville, IN 46060 USA Copyright © 2017 Helmer, Inc. 360173-A/C