Инструкция на Течеискатель хладагентов Inficon D-TEK Select
Электронный течеискатель хладагентов Inficon D-TEK Select: подготовка к работе, поиск утечек, замена сенсора, техническое обслуживание. Руководство по эксплуатации. Документ на английском языке.
Загружаем документ…
Текст инструкции
-
DECLARATION OF CONFORMITY
This is to certify that this equipment, designed and manufactured by INFICON® Inc., Two Technology Place, East Syracuse, NY 13057 USA, meets the essential safety requirements of the European Union and is placed on the market accordingly. It has been constructed in accordance with good engineering…
-
Filter
Cartridge High / Low Indicators Tip On / Off Switch Charge Indicator Probe Cap Green / Yellow LEDs Power Jack Flexible Probe High / Low Sensitivity Switch IR Cell Access Door Rubberized Grip Headphone Jack Power Cell Compartment Getting Started The D-TEK Select is shipped with the infrared cell and…
-
Using Your INFICON D-TEK Select
WARNING DO NOT OPERATE THIS INSTRUMENT IN THE PRESENCE OF GASOLINE, NATURAL GAS, PROPANE, OR IN OTHER COMBUSTIVE ATMOSPHERES. Using the D-TEK Select is simple. Press the power switch once to turn the detector ON. The green ON indicator lamp will illuminate, and the yellow indicator lamps will…
-
4. When the D-TEK Select signals a leak, pull the probe away from the leak for a
moment, then bring it back to pinpoint the location. 5. If the concentration of the refrigerant gas is high, press the sensitivity switch to change to the LOW sensitivity setting. The LOW sensitivity setting helps find the exact site when a leak is large. 6. Once you have isolated the leak source…
-
DO NOT ATTACH ANY OTHER POWER CORDS TO THE D-TEK
SELECT’S POWER JACK EXCEPT THE AC POWER ADAPTER AND DC CORDS DESIGNED FOR THIS INSTRUMENT. The battery charging indicator (rear left side) will flash when the power stick is charging, changing to steady illumination when the power stick is fully charged. Allow 10 hours for a full charge. The D-TEK…
-
Changing the Infrared Cell
WARNING COMPONENTS MAY BE HOT. TURN OFF DETECTOR AND DISCONNECT POWER ADAPTER CORD BEFORE REMOVING THE INFRARED CELL DOOR. The D-TEK Select’s infrared cell (IR cell) is located in the body of the detector. The IR cell is a complete assembly consisting of a metal tube, connectors and electronic…
-
NOTE
If the D-TEK Select does not power up after the power stick is replaced, check that the key connector on the power stick is mated properly to the connector on the circuit board. Reverse the connection if necessary and power up again. If the Hi/Lo indicators flash in a rapid, alternating manner when…
-
Troubleshooting Guide
Problem Cause Remedy 1) All yellow lights flashing together. 1a) IR sensor cell has become unseated. 1a) Remove sensor access door and push both ends of sensor down. (Do not remove/reinsert sensor cell.) Restart unit and examine. 1b) IR sensor cell has failed. 1b) Replace with new sensor, part…
-
Problem
Cause Remedy 3) After warm up sequence the green, the last yellow and the amber charging LEDs flash. 3a) Battery needs recharging. 3a) Charge battery for 10-12 hours. 3b) Power stick has failed. 3b) Replace power stick with part #712-700-G1. 4) Pump is not working. 4a) Pump has failed. 4a) See 2b.…
-
Return Materials Authorization Procedure
All instruments and parts returned to INFICON for repair or credit must be properly packaged, insured, shipped transportation charges prepaid, and must have a Return Material Authorization (RMA) number issued before the material is returned. The RMA number is to be marked on all shipping labels and…
-
TWO TECHNOLOGY PLACE
EAST SYRACUSE, NY 13057-9714 USA Phone: +315.434.1100 Fax: +315.437.3803 E-Mail: service.tools@inficon.com www.inficon.com 074-392-P1H